Pure Chromatography
www.restek.com
www.restek.com

This chromatogram may contain analytes that are synonyms of your search terms.
Scotch on Rxi-1301Sil MS
Peaks | tR (min) | |
---|---|---|
1. | Acetaldehyde![]() | 4.409 |
2. | Methanol![]() | 4.511 |
3. | Ethanol![]() | 5.440 |
4. | n-Propanol![]() | 6.883 |
Peaks | tR (min) | |
---|---|---|
5. | Ethyl acetate![]() | 7.452 |
6. | Isobutanol![]() | 8.182 |
7. | Acetal![]() | 9.509 |
8. | Isoamyl alcohol (3-methyl-1-butanol)![]() | 10.361 |
9. | Active amyl alcohol (2-methyl-1-butanol)![]() | 10.432 |

GC_FF1277
Column | Rxi-1301Sil MS, 60 m, 0.25 mm ID, 1.00 µm (cat.# 16097) | ||||||||
Sample | Scotch | ||||||||
Conc.: | Neat | ||||||||
Injection | |||||||||
Inj. Vol.: | 1 µL split (split ratio 20:1) | ||||||||
Liner: | Topaz 4.0 mm ID Precision inlet liner w/wool (cat.# 23305) | ||||||||
Inj. Temp.: | 250 °C | ||||||||
Oven | |||||||||
Oven Temp.: | 40 °C (hold 1 min) to 300 °C at 10 °C/min (hold 1 min) | ||||||||
Carrier Gas | He, constant flow | ||||||||
Flow Rate: | 1.4 mL/min | ||||||||
Detector | MS | ||||||||
Mode: | Scan | ||||||||
Scan Program: |
| ||||||||
Transfer Line Temp.: | 280 °C | ||||||||
Analyzer Type: | Quadrupole | ||||||||
Source Type: | Inert | ||||||||
Source Temp.: | 230 °C | ||||||||
Quad Temp.: | 150 °C | ||||||||
Electron Energy: | 70 eV | ||||||||
Tune Type: | BFB | ||||||||
Ionization Mode: | EI | ||||||||
Instrument | Agilent 7890A GC & 5975C MSD | ||||||||
Notes | Water coelutes with acetaldehyde and methanol, which results in distortion of the baseline. |
RELATED CHROMATOGRAM LIBRARY SEARCHES
Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.