Scotch on Rxi-1301Sil MS
This chromatogram may contain analytes that are synonyms of your search terms.

Scotch on Rxi-1301Sil MS

PeakstR (min)
1.Acetaldehyde4.409
2.Methanol4.511
3.Ethanol5.440
4.n-Propanol6.883
PeakstR (min)
5.Ethyl acetate7.452
6.Isobutanol8.182
7.Acetal9.509
8.Isoamyl alcohol (3-methyl-1-butanol)10.361
9.Active amyl alcohol (2-methyl-1-butanol)10.432
image
GC_FF1277
ColumnRxi-1301Sil MS, 60 m, 0.25 mm ID, 1.00 µm (cat.# 16097)
SampleScotch
Conc.:Neat
Injection
Inj. Vol.:1 µL split (split ratio 20:1)
Liner:Topaz 4.0 mm ID Precision inlet liner w/wool (cat.# 23305)
Inj. Temp.:250 °C
Oven
Oven Temp.:40 °C (hold 1 min) to 300 °C at 10 °C/min (hold 1 min)
Carrier GasHe, constant flow
Flow Rate:1.4 mL/min
DetectorMS
Mode:Scan
Scan Program:
GroupStart Time
(min)
Scan Range
(amu)
Scan Rate
(scans/sec)
1420-3002.5
Transfer Line Temp.:280 °C
Analyzer Type:Quadrupole
Source Type:Inert
Source Temp.:230 °C
Quad Temp.:150 °C
Electron Energy:70 eV
Tune Type:BFB
Ionization Mode:EI
InstrumentAgilent 7890A GC & 5975C MSD
NotesWater coelutes with acetaldehyde and methanol, which results in distortion of the baseline.
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

fusel alcohol

,

liquor analysis

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)
Drag and Drop Attachments Here (optional)
...


all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us