Rxi-65TG: High-Temperature Stability Overlay
This chromatogram may contain analytes that are synonyms of your search terms.

Rxi-65TG: High-Temperature Stability Overlay

Peaks
1.Tripalmitin (PPP)
2.1,2-Palmitin-3-stearin (PPS)
3.1,3-Palmitin-2-olein (POP)
4.1,2-Palmitin-3-linolein (PPL)
5.1-Palmitin-2-olein-3-stearin (POS)
Peaks
6.1,2-Olein-3-palmitin (POO)
7.1-Palmitin-2-linolein-3-olein (PLO)
8.1,2-Stearin-3-olein (SSO)
9.Triolein (OOO)
10.1,2-Olein-3-linolein (OOL)
image
GC_FF1287
ColumnRxi-65TG, 30 m, 0.25 mm ID, 0.1 µm (cat.# 17108)
SamplePalm oil analytical standard
Mono/diacylglycerol custom mix
Diluent:Isooctane
Injection
Inj. Vol.:1 µL split (split ratio 100:1)
Liner:Topaz 4.0 mm ID Precision inlet liner w/wool (cat.# 23305)
Inj. Temp.:360 °C
Split Vent Flow Rate:200 mL/min
Oven
Carrier GasH2, constant flow
Flow Rate:2 mL/min
DetectorFID @ 360 °C
Make-up Gas Flow Rate:30 mL/min
Make-up Gas Type:N2
Hydrogen flow:40 mL/min
Air flow:370 mL/min
Data Rate:50 Hz
InstrumentAgilent 7890B GC
NotesSample prep:
~50 mg of a palm oil standard (which contains naturally occurring triglycerides) was fortified with the following mono- and diacylglycerol standards. The mixture was then brought up to 10 mL in isooctane, giving a final mono- and diglycerides concentration of ~100 μg/mL.
1. 1,2-Dipalmitin
2. 1,3-Dipalmitin
3. 1,2-Distearin
4. 1,3-Distearin
5. 1-Monopalmitin
6. 2-Monopalmitin
7. 1-Monostearin

Testing Conditions:
Six injections of fortified palm oil were made under the following conditions. Extended final hold times were used to challenge column stability. After each 24-hour challenge period, column performance was assessed using an injection with a final hold time that is typical for triglyceride analysis.

Injection 1: 100 °C (hold 1 min) to 325 °C at 30 °C/min to 345 °C at 1 °C/min to 360 °C at 5 °C/min (hold 12 hr)
Injection 2: 100 °C (hold 1 min) to 325 °C at 30 °C/min to 345 °C at 1 °C/min to 360 °C at 5 °C/min (hold 12 hr)
Injection 3: 100 °C (hold 1 min) to 325 °C at 30 °C/min to 345 °C at 1 °C/min to 360 °C at 5 °C/min (hold 5 min)
Injection 4: 100 °C (hold 1 min) to 325 °C at 30 °C/min to 345 °C at 1 °C/min to 360 °C at 5 °C/min (hold 12 hr)
Injection 5: 100 °C (hold 1 min) to 325 °C at 30 °C/min to 345 °C at 1 °C/min to 360 °C at 5 °C/min (hold 12 hr)
Injection 6: 100 °C (hold 1 min) to 325 °C at 30 °C/min to 345 °C at 1 °C/min to 360 °C at 5 °C/min (hold 5 min)
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Triacylglycerols

,

Thermal Stability

,

Rxi-65TG

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)
Drag and Drop Attachments Here (optional)
...


all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us