share:
delim
delim
Extra Virgin Olive Oil on Rxi-65TG
This chromatogram may contain analytes that are synonyms of your search terms.

Extra Virgin Olive Oil on Rxi-65TG

PeakstR (min)
1.1,3-Palmitin-2-olein (POP)16.196
2.1,2-Palmitin-3-linolein (PPL)16.530
3.1-Palmitin-2-olein-3-stearin (POS)18.287
4.1,2-Olein-3-palmitin (POO)18.687
PeakstR (min)
5.1-Palmitin-2-linolein-3-olein (PLO)19.133
6.1,2-Linolein-3-palmitin (PLL)19.593
7.1,2-Stearin-3-olein (SOS)20.747
8.Triolein (OOO)21.612
image
GC_FF1293
ColumnRxi-65TG, 30 m, 0.25 mm ID, 0.1 µm (cat.# 17108)
SampleExtra virgin olive oil
Conc.: ~50 mg of extra virgin olive oil was diluted to 10 mL with isooctane.
Injection
Inj. Vol.:1 µL split (split ratio 35:1)
Liner:Topaz 4.0 mm ID Precision inlet liner w/wool (cat.# 23305)
Inj. Temp.:360 °C
Split Vent Flow Rate:35 mL/min
Oven
Oven Temp.:200 °C (hold 1 min) to 350 °C at 20 °C/min to 365 °C at 1 °C/min (hold 6 min)
Carrier GasH2, constant flow
Flow Rate:1 mL/min
DetectorFID @ 365 °C
Make-up Gas Flow Rate:30 mL/min
Make-up Gas Type:N2
Hydrogen flow:40 mL/min
Air flow:370 mL/min
Data Rate:50 Hz
InstrumentAgilent 7890B GC
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

extra virgin olive oil

,

triglyceride analysis

,

Rxi-65TG

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)
Drag and Drop Attachments Here (optional)
...


all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us