Tea Tree Oil on Rtx-Wax
Peaks | tR (min) | |
---|---|---|
1. | α-Pinene![]() | 1.473 |
2. | Terpinolene![]() | 1.757 |
3. | D-Limonene![]() | 1.806 |
4. | Eucalyptol![]() | 1.841 |
5. | γ-Terpinene![]() | 1.935 |
6. | p-Cymene![]() | 2.014 |
7. | γ-Elemene![]() | 2.953 |
8. | Terpinen-4-ol![]() | 3.570 |
9. | Caryophyllene![]() | 3.639 |

GC_FF1304
Column | Rtx-Wax, 30 m, 0.32 mm ID, 0.25 µm (cat.# 12424) |
Sample | Tea tree oil |
Diluent: | Acetone |
Conc.: | 5% |
Injection | |
Inj. Vol.: | 1 µL split (split ratio 100:1) |
Liner: | Topaz 4.0 mm ID Precision inlet liner w/wool (cat.# 23305) |
Inj. Temp.: | 230 °C |
Oven | |
Oven Temp.: | 100 °C (hold 0.5 min) to 250 °C at 16 °C/min (hold 10 min) |
Carrier Gas | H2, constant flow |
Flow Rate: | 2 mL/min |
Detector | FID @ 250 °C |
Constant Column + Constant Make-up: | 52 mL/min |
Make-up Gas Type: | N2 |
Hydrogen flow: | 40 mL/min |
Air flow: | 400 mL/min |
Data Rate: | 50 Hz |
Instrument | Agilent 7890A GC |
Notes | All peaks were identified using the NIST MS EI spectra library (2005). |