Tea Tree Oil on Rtx-Wax

PeakstR (min)
1.α-Pinene1.473
2.Terpinolene1.757
3.D-Limonene1.806
4.Eucalyptol1.841
5.γ-Terpinene1.935
6.p-Cymene2.014
7.γ-Elemene2.953
8.Terpinen-4-ol3.570
9.Caryophyllene3.639
PeakstR (min)
10.Aromadendrene3.697
11.Pinocarveol3.880
12.Alloaromadendrene3.920
13.α-Caryophyllene4.054
14.α-Terpineol4.098
15.Bornyl acetate4.148
16.γ-Elemene4.202
17.Nerol4.716
18.Geraniol4.993
19.Safrole5.223
image
GC_FF1304
ColumnRtx-Wax, 30 m, 0.32 mm ID, 0.25 µm (cat.# 12424)
SampleTea tree oil
Diluent:Acetone
Conc.:5%
Injection
Inj. Vol.:1 µL split (split ratio 100:1)
Liner:Topaz 4.0 mm ID Precision inlet liner w/wool (cat.# 23305)
Inj. Temp.:230 °C
Oven
Oven Temp.:100 °C (hold 0.5 min) to 250 °C at 16 °C/min (hold 10 min)
Carrier GasH2, constant flow
Flow Rate:2 mL/min
DetectorFID @ 250 °C
Constant Column + Constant Make-up:52 mL/min
Make-up Gas Type:N2
Hydrogen flow:40 mL/min
Air flow:400 mL/min
Data Rate:50 Hz
InstrumentAgilent 7890A GC
NotesAll peaks were identified using the NIST MS EI spectra library (2005).
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

terpinene

,

terpineol

,

essential oils

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us