Citronella Oil on Rxi-5Sil MS
This chromatogram may contain analytes that are synonyms of your search terms.

Citronella Oil on Rxi-5Sil MS

PeakstR (min)
1.α-Pinene2.059
2.D-Limonene2.545
3.Linalool2.958
4.Citronellal3.400
5.Isopulegol isomer 13.455
6.Isopulegol isomer 23.549
7.α-Terpineol3.844
8.β-Citronellol4.042
9.Z-Citral4.191
10.Geraniol4.292
11.E-Citral4.462
12.Citronellyl propionate5.257
13.Eugenol5.371
14.Geranyl acetate5.558
PeakstR (min)
15.β-Elemene5.822
16.Caryophyllene6.212
17.α-Caryophyllene6.587
18.α-Amorphene6.743
19.Germacrene D6.845
20.α-Muurolene6.985
21.δ-Cadinene7.189
22.α-Cadinene7.398
23.Elemol7.495
24.Germacrene D7.835
25.γ-Eudesmol8.400
26.α-Cadinol8.501
27.α-Eudesmol8.626
28.β-Eudesmol8.656
image
GC_FF1308
ColumnRxi-5Sil MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13623)
SampleCitronella oil
Diluent:Acetone
Conc.:5%
Injection
Inj. Vol.:1 µL split (split ratio 100:1)
Liner:Topaz 4.0 mm ID Precision inlet liner w/wool (cat.# 23305)
Inj. Temp.:250 °C
Oven
Oven Temp.:100 °C to 300 °C at 11 °C/min (hold 10 min)
Carrier GasHe, constant flow
Flow Rate:1.31 mL/min
DetectorMS
Mode:Scan
Scan Program:
GroupStart Time
(min)
Scan Range
(amu)
Scan Rate
(scans/sec)
11.0035-5005
Transfer Line Temp.:300 °C
Analyzer Type:Quadrupole
Source Type:Inert
Source Temp.:230 °C
Quad Temp.:150 °C
InstrumentAgilent 7890A GC & 5975C MSD
NotesAll peaks were identified using the NIST MS EI spectra library (2005).
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Essential oils

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us