Eucalyptus Oil on Rxi-5Sil MS
Peaks | tR (min) | |
---|---|---|
1. | α-Pinene![]() | 2.066 |
2. | β-Myrcene![]() | 2.155 |
3. | β-Pinene![]() | 2.254 |
4. | α-Phellandrene![]() | 2.290 |
5. | Terpinolene![]() | 2.414 |
6. | p-Cymene![]() | 2.471 |
7. | D-Limonene![]() | 2.558 |
8. | Eucalyptol![]() | 2.602 |
9. | γ-Terpinene![]() | 2.720 |

GC_FF1314
Column | Rxi-5Sil MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13623) | ||||||||
Sample | Eucalyptus oil | ||||||||
Diluent: | Acetone | ||||||||
Conc.: | 5% | ||||||||
Injection | |||||||||
Inj. Vol.: | 1 µL split (split ratio 100:1) | ||||||||
Liner: | Topaz 4.0 mm ID Precision inlet liner w/wool (cat.# 23305) | ||||||||
Inj. Temp.: | 250 °C | ||||||||
Oven | |||||||||
Oven Temp.: | 100 °C to 300 °C at 11 °C/min (hold 10 min) | ||||||||
Carrier Gas | He, constant flow | ||||||||
Flow Rate: | 1.31 mL/min | ||||||||
Detector | MS | ||||||||
Mode: | Scan | ||||||||
Scan Program: |
| ||||||||
Transfer Line Temp.: | 300 °C | ||||||||
Analyzer Type: | Quadrupole | ||||||||
Source Type: | Inert | ||||||||
Source Temp.: | 230 °C | ||||||||
Quad Temp.: | 150 °C | ||||||||
Instrument | Agilent 7890A GC & 5975C MSD | ||||||||
Notes | All peaks were identified using the NIST MS EI spectra library (2005). |