share:
delim
delim
Geranium Oil on Rxi-5Sil MS

Geranium Oil on Rxi-5Sil MS

PeakstR (min)
1.Linalool2.964
2.Rose oxide 13.074
3.Rose oxide 23.216
4.Menthone3.509
5.Isomenthone3.596
6.α-Terpineol3.838
7.Citronellol4.054
8.Z-Citral4.192
9.Geraniol4.289
10.Citronellyl formate4.500
11.Geranyl formate4.754
12.cis-2,6-Dimethyl-2,6-octadiene5.254
13.α-Cubebene5.373
14.Geranyl acetate5.557
PeakstR (min)
15.Copaene5.714
16.β-Bourbonene5.822
17.Caryophyllene6.219
18.Neryl acetate6.529
19.Germacrene D6.850
20.γ-Gurjunene6.946
21.α-Muurolene7.000
22.Citronellyl butyrate7.135
23.δ-Cadinene7.189
24.Calamenene7.238
25.Geranyl butyrate7.457
26.Phenylethyl butyrate7.810
27.γ-Eudesmol8.351
28.Phenylethyl tiglate8.910
image
GC_FF1318
ColumnRxi-5Sil MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13623)
SampleGeranium oil
Diluent:Acetone
Conc.:5%
Injection
Inj. Vol.:1 µL split (split ratio 100:1)
Liner:Topaz 4.0 mm ID Precision inlet liner w/wool (cat.# 23305)
Inj. Temp.:250 °C
Oven
Oven Temp.:100 °C to 300 °C at 11 °C/min (hold 10 min)
Carrier GasHe, constant flow
Flow Rate:1.31 mL/min
DetectorMS
Mode:Scan
Scan Program:
GroupStart Time
(min)
Scan Range
(amu)
Scan Rate
(scans/sec)
11.0035-5005
Transfer Line Temp.:300 °C
Analyzer Type:Quadrupole
Source Type:Inert
Source Temp.:230 °C
Quad Temp.:150 °C
InstrumentAgilent 7890A GC & 5975C MSD
NotesAll peaks were identified using the NIST MS EI spectra library (2005).
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Essential oils

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)
Drag and Drop Attachments Here (optional)
...


all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us