Bergamot Oil on Rxi-5Sil MS (20 m, 0.18 mm ID, 0.18 µm)
Peaks | tR (min) | |
---|---|---|
1. | α-Thujene![]() | 1.212 |
2. | α-Pinene![]() | 1.253 |
3. | Sabinene![]() | 1.362 |
4. | β-Pinene![]() | 1.395 |
5. | p-Cymene![]() | 1.549 |
6. | D-Limonene![]() | 1.578 |
7. | γ-Terpinene![]() | 1.696 |
8. | Terpinolene![]() | 1.818 |
9. | Linalool![]() | 1.858 |

GC_FF1321
Column | Rxi-5Sil MS, 20 m, 0.18 mm ID, 0.18 µm (cat.# 43602) | ||||||||
Sample | Bergamot oil | ||||||||
Diluent: | Acetone | ||||||||
Conc.: | 5% | ||||||||
Injection | |||||||||
Inj. Vol.: | 1 µL split (split ratio 100:1) | ||||||||
Liner: | Topaz 4.0 mm ID Precision inlet liner w/wool (cat.# 23305) | ||||||||
Inj. Temp.: | 250 °C | ||||||||
Oven | |||||||||
Oven Temp.: | 100 °C (hold 0.25 min) to 320 °C at 17.5 °C/min (hold 10 min) | ||||||||
Carrier Gas | He, constant flow | ||||||||
Flow Rate: | 1.01 mL/min | ||||||||
Detector | MS | ||||||||
Mode: | Scan | ||||||||
Scan Program: |
| ||||||||
Transfer Line Temp.: | 300 °C | ||||||||
Analyzer Type: | Quadrupole | ||||||||
Source Type: | Inert | ||||||||
Source Temp.: | 230 °C | ||||||||
Quad Temp.: | 150 °C | ||||||||
Instrument | Agilent 7890A GC & 5975C MSD | ||||||||
Notes | All peaks were identified using the NIST MS EI spectra library (2005). |