delim
delim
Patchouli Oil on Rxi-5Sil MS (20 m, 0.18 mm ID, 0.18 µm)
This chromatogram may contain analytes that are synonyms of your search terms.

Patchouli Oil on Rxi-5Sil MS (20 m, 0.18 mm ID, 0.18 µm)

PeakstR (min)
1.α-Pinene1.248
2.β-Pinene1.399
3.D-Limonene1.569
4.Eucalyptol1.590
5.δ-Elemene3.331
6.β-Elemene3.704
7.β-Patchoulene3.725
8.Caryophyllene3.955
PeakstR (min)
9.α-Guaiene4.047
10.Seychellene4.198
11.α-Patchoulene4.275
12.Patchoulene4.318
13.Cedrene4.450
14.δ-Guaiene4.509
15.Valencene5.553
16.Patchoulol5.700
image
GC_FF1324
ColumnRxi-5Sil MS, 20 m, 0.18 mm ID, 0.18 µm (cat.# 43602)
SamplePatchouli oil
Diluent:Acetone
Conc.:5%
Injection
Inj. Vol.:1 µL split (split ratio 100:1)
Liner:Topaz 4.0 mm ID Precision inlet liner w/wool (cat.# 23305)
Inj. Temp.:250 °C
Oven
Oven Temp.:100 °C (hold 0.25 min) to 320 °C at 17.5 °C/min (hold 10 min)
Carrier GasHe, constant flow
Flow Rate:1.01 mL/min
DetectorMS
Mode:Scan
Scan Program:
GroupStart Time
(min)
Scan Range
(amu)
Scan Rate
(scans/sec)
11.0035-5005
Transfer Line Temp.:300 °C
Analyzer Type:Quadrupole
Source Type:Inert
Source Temp.:230 °C
Quad Temp.:150 °C
InstrumentAgilent 7890A GC & 5975C MSD
NotesAll peaks were identified using the NIST MS EI spectra library (2005).
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Essential oils

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us