share:
delim
delim
Geranium Oil on Rxi-5Sil MS (10 m, 0.15 mm ID, 0.15 µm)
This chromatogram may contain analytes that are synonyms of your search terms.

Geranium Oil on Rxi-5Sil MS (10 m, 0.15 mm ID, 0.15 µm)

PeakstR (min)
1.Linalool0.732
2.Rose oxide 10.756
3.Rose oxide 20.791
4.Menthone0.862
5.Isomenthone0.885
6.α-Terpineol0.943
7.Citronellol1.010
8.Z-Citral1.035
9.Geraniol1.065
10.Citronellyl formate1.109
11.Geranyl formate1.167
12.2,6-Dimethyl-2,6-octadiene1.283
13.α-Cubebene1.312
14.Geranyl acetate1.357
PeakstR (min)
15.Copaene1.394
16.β-Bourbonene1.420
17.Caryophyllene1.516
18.Neryl acetate1.593
19.Germacrene D1.670
20.γ-Gurjunene1.694
21.α-Muurolene1.705
22.Citronellyl butyrate1.741
23.δ-Cadinene1.754
24.Calamenene1.766
25.Geranyl butyrate1.818
26.Phenylethyl butyrate1.906
27.γ-Eudesmol2.042
28.Phenylethyl tiglate2.175
image
GC_FF1340
ColumnRxi-5Sil MS, 10 m, 0.15 mm ID, 0.15 µm (cat.# 43815)
SampleGeranium oil
Diluent:Acetone
Conc.:1%
Injection
Inj. Vol.:1 µL split (split ratio 100:1)
Liner:Topaz 4.0 mm ID Precision inlet liner w/wool (cat.# 23305)
Inj. Temp.:250 °C
Oven
Oven Temp.:100 °C to 300 °C at 45 °C/min to 320 °C at 30 °C/min (hold 5 min)
Carrier GasHe, constant flow
Flow Rate:1.01 mL/min
DetectorMS
Mode:Scan
Scan Program:
GroupStart Time
(min)
Scan Range
(amu)
Scan Rate
(scans/sec)
11.0035-50011
Transfer Line Temp.:300 °C
Analyzer Type:Quadrupole
Source Type:Inert
Source Temp.:230 °C
Quad Temp.:150 °C
InstrumentAgilent 7890A GC & 5975C MSD
NotesAll peaks were identified using the NIST MS EI spectra library (2005).
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Essential oils

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us