share:
delim
delim
Lemongrass Oil on Rxi-5Sil MS (10 m, 0.15 mm ID, 0.15 µm)

Lemongrass Oil on Rxi-5Sil MS (10 m, 0.15 mm ID, 0.15 µm)

PeakstR (min)
1.Camphene0.536
2.5-Hepten-2-one, 6-methyl-0.551
3.β-Ocimene0.620
4.D-Limonene0.629
5.α-Ocimene0.639
6.4-Nonanone0.677
7.Linalool0.729
PeakstR (min)
8.Z-Citral1.038
9.Geraniol1.053
10.E-Citral1.111
11.Geranyl acetate1.358
12.Caryophyllene1.514
13.α-Caryophyllene1.556
14.α-Amorphene1.744
15.δ-Cadinene1.751
image
GC_FF1341
ColumnRxi-5Sil MS, 10 m, 0.15 mm ID, 0.15 µm (cat.# 43815)
SampleLemongrass oil
Diluent:Acetone
Conc.:1%
Injection
Inj. Vol.:1 µL split (split ratio 100:1)
Liner:Topaz 4.0 mm ID Precision inlet liner w/wool (cat.# 23305)
Inj. Temp.:250 °C
Oven
Oven Temp.:100 °C to 300 °C at 45 °C/min to 320 °C at 30 °C/min (hold 5 min)
Carrier GasHe, constant flow
Flow Rate:1.01 mL/min
DetectorMS
Mode:Scan
Scan Program:
GroupStart Time
(min)
Scan Range
(amu)
Scan Rate
(scans/sec)
11.0035-50011
Transfer Line Temp.:300 °C
Analyzer Type:Quadrupole
Source Type:Inert
Source Temp.:230 °C
Quad Temp.:150 °C
InstrumentAgilent 7890A GC & 5975C MSD
NotesAll peaks were identified using the NIST MS EI spectra library (2005).
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Essential oils

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)
Drag and Drop Attachments Here (optional)
...


all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us