share:
delim
delim
Palm Oil on Rxi-65TG (30 m x 0.32 mm x 0.1 µm)
This chromatogram may contain analytes that are synonyms of your search terms.

Palm Oil on Rxi-65TG (30 m x 0.32 mm x 0.1 µm)

PeakstR (min)
1.Tripalmitin (PPP)12.988
2.1,2-Palmitin-3-stearin (PPS)14.395
3.1,3-Palmitin-2-olein (POP)14.640
4.1,2-Palmitin-3-linolein (PPL)14.953
5.1-Palmitin-2-olein-3-stearin (POS)16.331
PeakstR (min)
6.1,2-Olein-3-palmitin (POO)16.656
7.1-Palmitin-2-linolein-3-olein (PLO)17.035
8.1,2-Linolein-3-palmitin (PLL)17.416
9.1,2-Stearin-3-olein (SOS)18.322
10.Triolein (OOO)18.940
image
GC_FF1342
ColumnRxi-65TG, 30 m, 0.32 mm ID, 0.10 µm (cat.# 17109)
SamplePalm oil
Diluent:Isoocatane
Conc.: ~5 mg/mL
Injection
Inj. Vol.:1 µL split (split ratio 25:1)
Liner:Topaz 4.0 mm ID Precision inlet liner w/wool (cat.# 23305)
Inj. Temp.:360 °C
Oven
Oven Temp.:200 °C (hold 1.2 min) to 350 °C at 21.7 °C/min to 365 °C at 1.2 °C/min (hold 5 min)
Carrier GasH2, constant flow
Flow Rate:1.28 mL/min
Dead Time:1.4331 min
DetectorFID @ 365 °C
Make-up Gas Flow Rate:30 mL/min
Make-up Gas Type:N2
Hydrogen flow:40 mL/min
Air flow:370 mL/min
Data Rate:50 Hz
InstrumentAgilent 7890B GC
NotesSample Preparation: ~50 mg of palm oil was diluted to 10 mL with isooctane.
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

edible oil analysis

,

palm oil analysis

,

Rxi-65TG

,

triglycerides

,

triglyceride analysis

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)
Drag and Drop Attachments Here (optional)
...


all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us