Pure Chromatography
www.restek.com
www.restek.com



Extra Virgin Olive Oil on Rxi-65TG (30 m x 0.32 mm x 0.1 µm)
Peaks | tR (min) | |
---|---|---|
1. | 1,3-Palmitin-2-olein (POP)![]() | 14.592 |
2. | 1,2-Palmitin-3-linolein (PPL)![]() | 14.875 |
3. | 1-Palmitin-2-olein-3-stearin (POS)![]() | 16.324 |
4. | 1,2-Olein-3-palmitin (POO) | 16.649 |
Peaks | tR (min) | |
---|---|---|
5. | 1-Palmitin-2-linolein-3-olein (PLO)![]() | 17.013 |
6. | 1,2-Linolein-3-palmitin (PLL) | 17.397 |
7. | 1,2-Stearin-3-olein (SOS)![]() | 18.348 |
8. | Triolein (OOO)![]() | 19.065 |

GC_FF1343
Column | Rxi-65TG, 30 m, 0.32 mm ID, 0.10 µm (cat.# 17109) |
Sample | Extra virgin olive oil |
Diluent: | Isooctane |
Conc.: | 5 mg/mL |
Injection | |
Inj. Vol.: | 1 µL split (split ratio 25:1) |
Liner: | Topaz 4.0 mm ID Precision inlet liner w/wool (cat.# 23305) |
Inj. Temp.: | 360 °C |
Oven | |
Oven Temp.: | 200 °C (hold 1.2 min) to 350 °C at 21.7 °C/min to 365 °C at 1.2 °C/min (hold 5 min) |
Carrier Gas | H2, constant flow |
Flow Rate: | 1.28 mL/min |
Dead Time: | 1.4331 min |
Detector | FID @ 365 °C |
Make-up Gas Flow Rate: | 30 mL/min |
Make-up Gas Type: | N2 |
Hydrogen flow: | 40 mL/min |
Air flow: | 370 mL/min |
Data Rate: | 50 Hz |
Instrument | Agilent 7890B GC |
Notes | Sample Preparation: 50 mg of extra virgin olive oil was diluted to 10 mL with isooctane. |
RELATED CHROMATOGRAM LIBRARY SEARCHES
Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.