Medical Cannabis Terpenes on Rxi-624Sil MS
This chromatogram may contain analytes that are synonyms of your search terms.

Medical Cannabis Terpenes on Rxi-624Sil MS

PeakstR (sec)
1.alpha-Pinene427.5
2.Camphene446.2
3.Myrcene465.7
4.beta-Pinene470.0
5.delta-3-Carene490.3
6.alpha-Terpinene497.7
7.Limonene506.6
8.1,8-Cineole518.3
9.gamma-Terpinene527.8
10.Terpinolene551.6
11.Linalool576.2
12.Fenchone587.0
13.Isopulegol627.0
14.Menthol647.9
PeakstR (sec)
15.Borneol654.6
16.alpha-Terpineol660.5
17.Dihydrocarveol667.3
18.Citronellol674.6
19.Geraniol693.6
20.2-Piperidone696.9
21.Citral 1698.7
22.Pulegone701.1
23.Citral 2717.7
24.Citral 3775.2
25.Citral 4789.4
26.beta-Caryophyllene812.2
27.Nerolidol 1870.2
28.Nerolidol 2888.4
29.Caryophyllene oxide937.1
image
GC_FS0509
ColumnRxi-624Sil MS, 30 m, 0.25 mm ID, 1.40 µm (cat.# 13868)
Injection
Inj. Vol.:1 µL split (split ratio 10:1)
Liner:Premium 4 mm Precision w/wool (cat.# 23305.5)
Inj. Temp.:250 °C
Oven
Oven Temp.:60 °C (hold 0.1 min) to 300 °C at 12.5 °C/min (hold 1.7 min)
Carrier GasHe, constant flow
Flow Rate:1.4 mL/min
DetectorFID @ 320 °C
Constant Column + Constant Make-up:50 mL/min
Make-up Gas Type:N2
Hydrogen flow:40 mL/min
Air flow:450 mL/min
Data Rate:10 Hz
InstrumentAgilent/HP6890 GC
AcknowledgementAndrew Goldsmith - SRI Instruments
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

terpenes

,

medical cannabis

,

medical marijuana

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)
Drag and Drop Attachments Here (optional)
...


all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us