share:
delim
delim
Cannabis Terpenes on Rxi-624Sil MS by FET-HS-GC
This chromatogram may contain analytes that are synonyms of your search terms.

Cannabis Terpenes on Rxi-624Sil MS by FET-HS-GC

PeakstR (min)
1.α-Pinene7.39
2.Camphene7.71
3.β-Myrcene7.98
4.Sabinene8.02
5.β-Pinene8.11
6.α-Phellandrene8.4
7.δ 3-Carene8.44
8.α-Terpinene8.57
9.Ocimene8.61
10.Limonene8.71
11.p-Cymene8.75
12.β-Ocimene8.82
13.Eucalyptol8.91
14.γ-Terpinene9.06
15.Terpinolene9.47
16.Linalool9.87
17.Fenchone10.06
18.Isopulegol10.73
PeakstR (min)
19.dl-Menthol11.08
20.Borneol11.19
21.α-Terpineol11.29
22.Dihydrocarveol11.40
23.Citronellol11.51
24.Geraniol11.82
25.2-Piperidinone11.88
26.Citral 111.92
27.Pulegone11.97
28.Citral 212.24
29.Citral 313.19
30.Citral 413.43
31.β-caryophyllene13.83
32.α-Humulene14.21
33.Nerolidol 114.78
34.Nerolidol 215.08
35.Caryophyllene oxide15.92
36.α-Bisabolol16.43
image
GC_FS0518
ColumnRxi-624Sil MS, 30 m, 0.25 mm ID, 1.40 µm (cat.# 13868)
SampleTerpenes mix
Diluent:Isopropyl alcohol
Conc.:200 ng/µL (0.02% wt/vol). The sample was prepared by placing 10 µL into the headspace vial.
Injectionheadspace-loop split (split ratio 10:1)
Liner:Premium 1.0 mm ID straight inlet liner (cat.# 23333.1)
Headspace-Loop
Inj. Port Temp.:250 °C
Instrument:Tekmar HT-3
Inj. Time:1.0 min
Transfer Line Temp.:160 °C
Valve Oven Temp.:160 °C
Needle Temp.:140 °C
Sample Temp.:140 °C
Sample Equil. Time:30.0 min
Vial Pressure:20 psi
Loop Pressure:15 psi
Oven
Oven Temp.:60 °C (hold 0.10 min) to 300 °C at 12.50 °C/min (hold 3.0 min)
Carrier GasHe, constant flow
Linear Velocity:33 cm/sec
DetectorFID @ 320 °C
Make-up Gas Flow Rate:45 mL/min
Make-up Gas Type:N2
Hydrogen flow:40 mL/min
Air flow:450 mL/min
Data Rate:20 Hz
InstrumentAgilent/HP6890 GC
NotesFor qualitative purposes only.
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Terpenes

,

Cannabis

,

Marijuana

,

624Sil

,

FET

,

,

headspace

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us