Fast Analysis of 3-MCPD and 2-MCPD on Rxi-17Sil MS

Fast Analysis of 3-MCPD and 2-MCPD on Rxi-17Sil MS

PeakstR (min)Conc.
(ng/mL)
1.3-MCPD-d5 PBA derivative4.143100
2.3-MCPD PBA derivative4.169200
3.2-MCPD-d5 PBA derivative4.388100
4.2-MCPD PBA derivative4.418100
image
GC_FS0569
ColumnRxi-17Sil MS, 20 m, 0.18 mm ID, 0.18 µm (cat.# 14102)
SampleSee notes
Diluent:Isooctane
Injection
Inj. Vol.:1 µL PTV split (split ratio 10:1)
Liner:Topaz 2.0 mm ID straight inlet liner w/wool (cat.# 23314)
Inlet Temp. Program:120 °C to 165 °C at 300 °C/min (hold 10 min) to 320 °C at 300 °C/min (hold 8 min)
Oven
Oven Temp.:120 °C (hold 0.5 min) to 200 °C at 18.5 °C/min to 330 °C at 35 °C/min
Carrier GasHe, constant flow
Flow Rate:1 mL/min
DetectorMS
Mode:SIM
SIM Program:147, 150, 196, 201 m/z, 50 ms dwell
Transfer Line Temp.:320 °C
Analyzer Type:Quadrupole
Source Type:Inert
Source Temp.:230 °C
Quad Temp.:150 °C
Ionization Mode:EI
InstrumentAgilent 7890A GC & 5975C MSD
NotesStandards were derivatized with 20 µL phenylboronic acid (saturated solution in diethyl ether), dried, and then reconstituted in 1 mL isooctane. Final concentrations are given in the peak table.
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Edible oils

,

Infant formula

,

acid-hydrolysed vegetable proteins

,

soy sauce

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)
Drag and Drop Attachments Here (optional)
...


all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us