Comparison of Incurred Fludioxonil in Various Parts of Orange on Rxi-5ms

Comparison of Incurred Fludioxonil in Various Parts of Orange on Rxi-5ms

PeakstR (min)Precursor IonProduct IonCollision Energy (CE)
1.Fludioxonil18.1624812730
image
GC_FS0570
ColumnRxi-5ms, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13423)
SampleWhole orange QuEChERS extract (318 ng/g)
Orange peel QuEChERS extract (594 ng/g)
Orange pulp QuEChERS extract (6.8 ng/g)
Diluent:Acetonitrile
Injection
Inj. Vol.:1 µL splitless (hold 0.5 min)
Liner:Topaz 4.0 mm ID single taper inlet liner w/wool (cat.# 23447)
Inj. Temp.:250 °C
Oven
Oven Temp.:90 °C (hold 1 min) to 330 °C at 8.5 °C/min (hold 5 min)
Carrier GasHe, constant flow
Flow Rate:1.4 mL/min
DetectorTSQ 8000
Transfer Line Temp.:290 °C
Analyzer Type:Quadrupole
Source Temp.:280 °C
Tune Type:PFTBA
Ionization Mode:EI
InstrumentThermo Scientific TSQ 8000 Triple Quadrupole GC-MS
NotesSample Preparation (adapted from EN and AOAC QuEChERS methods):

Orange (Whole): 10 g of homogenized orange was fortified with an internal standard and mixed with 10 mL acetonitrile and two steel ball bearings (1/16'') and shaken for 10 minutes. Q-sep QuEChERS extraction salts for EN 15662 (cat.# 25849) were added, and the sample was shaken for 1 minute. The sample was centrifuged for 5 minutes, and the supernatant was removed and further cleaned up with prefilled Q-sep QuEChERS dSPE tubes containing 150 mg MgSO4, 50 mg PSA, and 50 mg C18-EC (cat.# 26125).

Orange Peel: 10 g of homogenized orange peel was fortified with an internal standard, hydrated with 4 mL water, and briefly shaken by hand. Acetonitrile (10 mL) was added, and the sample was shaken for an additional 1 minute. Q-sep QuEChERS extraction salts for EN 15662 (cat.# 25849) were added, and the sample was shaken for 1 minute. The sample was centrifuged for 5 minutes, and the supernatant was removed and further cleaned up with prefilled Q-sep QuEChERS dSPE tubes containing 150 mg MgSO4, 25 mg PSA, and 25 mg C18-EC (cat.# 26216).

Orange Pulp: 15 g of homogenized orange pulp was fortified with an internal standard. Acetonitrile with 1% acetic acid (15 mL) was added, and the sample was shaken for 1 minute. Q-sep QuEChERS extraction salts for AOAC 2007.01 (cat.# 25851) were added, and the sample was shaken for an additional 1 minute. The sample was centrifuged for 5 minutes, and the supernatant was removed and further cleaned up with prefilled Q-sep QuEChERS dSPE tubes containing 150 mg MgSO4 and 50 mg PSA (cat.# 26124).

The dSPE supernatant of all samples was analyzed directly (no further dilution) within 24 hours of the initial extraction.
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

QuEChERS

,

Pesticides

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)
Drag and Drop Attachments Here (optional)
...


all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us