GC Multiresidue Pesticide Standard #9-OPP on Rxi-5ms by GC-MS
This chromatogram may contain analytes that are synonyms of your search terms.

GC Multiresidue Pesticide Standard #9-OPP on Rxi-5ms by GC-MS

PeakstR (min)
1.Mevinphos8.92
2.Phorate12.59
3.Fonofos13.73
4.Disulfoton14.05
PeakstR (min)
5.Methyl parathion15.06
6.Parathion (ethyl parathion)16.22
7.Triazophos19.59
8.Piperonyl butoxide20.50
image
GC_FS0608
ColumnRxi-5ms, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13423)
SampleGC multiresidue pesticide standard #9-OPP (cat.# 32571)
Diluent:Toluene
Conc.:100 µg/mL
Injection
Inj. Vol.:1 µL split (split ratio 50:1)
Liner:Premium 4.0 mm ID Precision inlet liner w/wool (cat.# 23305.1)
Inj. Temp.:250 °C
Oven
Oven Temp.:90 °C (hold 1 min) to 330 °C at 8.5 °C/min (hold 5 min)
Carrier GasHe, constant flow
Flow Rate:1.4 mL/min
DetectorMS
Mode:Scan
Scan Program:
GroupStart Time
(min)
Scan Range
(amu)
Scan Rate
(scans/sec)
1555-5507
Transfer Line Temp.:290 °C
Analyzer Type:Quadrupole
Source Temp.:325 °C
Electron Energy:70 eV
Solvent Delay Time:5 min
Ionization Mode:EI
InstrumentThermo Scientific TSQ 8000 Triple Quadrupole GC-MS
NotesReconstructed chromatogram from select ions.
Download PDF

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)
Drag and Drop Attachments Here (optional)
...


all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us