QuEChERS Extracts of Tylenol Spiked with Anisoles and TBA on Rxi-5Sil MS

GC_GN1152
Column | Rxi-5Sil MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13623) |
Sample | QuEChERS extracts of Tylenol spiked with TBA and anisoles |
Diluent: | Acetonitrile |
Injection | |
Inj. Vol.: | 1 µL splitless (hold 1 min) |
Liner: | Gooseneck Splitless (4mm) w/Wool (cat.# 22405) |
Inj. Temp.: | 250 °C |
Purge Flow: | 40 mL/min |
Oven | |
Oven Temp.: | 90 °C (hold 1 min) to 310 °C at 4 °C/min (hold 2 min) |
Carrier Gas | He, constant flow |
Flow Rate: | 1.2 mL/min |
Linear Velocity: | 40 cm/sec |
Detector | TOFMS |
Transfer Line Temp.: | 290 °C |
Analyzer Type: | TOF |
Source Temp.: | 225 °C |
Electron Energy: | 70 eV |
Mass Defect: | -20 mu/100 u |
Solvent Delay Time: | 2 min |
Tune Type: | PFTBA |
Ionization Mode: | EI |
Acquisition Range: | 45 to 550 amu |
Spectral Acquisition Rate: | 5 spectra/sec |
Instrument | LECO Pegasus 4D GCxGC-TOFMS |
Notes | Total ion chromatogram (pink) of QuEChERS extracts of Tylenol spiked with anisoles and TBA, and extracted ion chromatogram (blue, 265x50) of pentachloroanisole, which was located by an automated peak find routine underneath the large acetaminophen peak. No dSPE was performed on this sample; see Applications Note GNAN1263 for complete sample fortification and extraction details. |