share:
delim
delim
Fragrance Allergens and Common Compounds in Commercial Perfume on Stabilwax-MS
This chromatogram may contain analytes that are synonyms of your search terms.

Fragrance Allergens and Common Compounds in Commercial Perfume on Stabilwax-MS

PeakstR (min)
1.Ethanol4.18
2.Linalool*18.59
3.Terpineol21.21
4.Benzyl acetate21.78
5.Citronellol*22.29
6.Dipropylene glycol isomer 123.31
7.Geraniol*23.57
PeakstR (min)
8.Dipropylene glycol isomer 224.10
9.Dipropylene glycol isomer 324.20
10.Lilial*26.63
11.Isoeugenol*28.29
12.Piperonal29.20
13.Diethyl phthalate30.91
14.Benzyl salicylate*35.81
* Denotes fragrance allergen or suspected allergen
image
GC_GN1161
ColumnStabilwax-MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 10673)
SampleCommercial perfume A
Conc.: Neat
Injection
Inj. Vol.:1 µL split (split ratio 200:1)
Liner:Premium 3.5 mm ID single taper w/wool (cat.# 23322.5)
Inj. Temp.:250 °C
Oven
Oven Temp.:35 °C (hold 5 min) to 250 °C at 7 °C/min (hold 5 min)
Carrier GasHe, constant linear velocity
Linear Velocity:36 cm/sec
DetectorMS
Mode:Scan
Scan Program:
GroupStart Time
(min)
Scan Range
(amu)
Scan Rate
(scans/sec)
1040-5502
Transfer Line Temp.:260 °C
Analyzer Type:Quadrupole
Source Temp.:250 °C
Electron Energy:70 eV
Solvent Delay Time:0.5 min
Ionization Mode:EI
InstrumentShimadzu 2010 GC & QP2010+ MS
Download PDF

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)
Drag and Drop Attachments Here (optional)
...


all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us