share:
delim
delim
Fragrance Allergens and Common Compounds in Commercial Perfume on Stabilwax-MS

Fragrance Allergens and Common Compounds in Commercial Perfume on Stabilwax-MS

PeakstR (min)
1.Ethanol4.18
2.Limonene*11.47
3.Linalool*18.58
4.Guaiene19.37
5.Citronellol*22.29
PeakstR (min)
6.Geraniol*23.57
7.Dipropylene glycol isomer 124.11
8.Dipropylene glycol isomer 224.21
9.Lilial*26.61
10.α-Hexyl-cinnamaldehyde*30.78
11.Benzyl salicylate*35.82
* Denotes fragrance allergen or suspected allergen; third isomer of dipropylene glycol not identified.
image
GC_GN1162
ColumnStabilwax-MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 10673)
SampleCommercial perfume B
Diluent:Neat
Injection
Inj. Vol.:1 µL split (split ratio 200:1)
Liner:Premium 3.5 mm single taper w/wool (cat.# 23322.5)
Inj. Temp.:250 °C
Oven
Oven Temp.:35 °C (hold 5 min) to 250 °C at 7 °C/min (hold 5 min)
Carrier GasHe, constant linear velocity
Linear Velocity:36 cm/sec
DetectorMS
Mode:Scan
Scan Program:
GroupStart Time
(min)
Scan Range
(amu)
Scan Rate
(scans/sec)
1040-5502
Transfer Line Temp.:260 °C
Analyzer Type:Quadrupole
Source Temp.:250 °C
Electron Energy:70 eV
Solvent Delay Time:0.5 min
Ionization Mode:EI
InstrumentShimadzu 2010 GC & QP2010+ MS
Download PDF

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Email

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us