share:
delim
delim
Custom Solvents Mix on Stabilwax-MS
This chromatogram may contain analytes that are synonyms of your search terms.

Custom Solvents Mix on Stabilwax-MS

Peaks
1.Ethyl acetate
2.n-Propyl acetate
3.MEK
4.Isopropyl alcohol
5.Ethanol
6.Isopropyl acetate
7.2-Pentanone
8.MIBK
9.1-Propanol
Peaks
10.Butyl acetate
11.1-Methoxy-2-propanol
12.p-Xylene
13.m-Xylene
14.Propylene glycol ethyl ether
15.o-Xylene
16.1-Methoxy-2-propyl acetate
17.Cyclohexanone
18.Ethylene glycol monoethyl ether acetate
19.Diacetone alcohol
image
GC_GN1163
ColumnStabilwax-MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 10673)
SampleCustom solvent standard
Diluent:Cyclohexane
Conc.:400 ppm
Injection
Inj. Vol.:0.5 µL split (split ratio 200:1)
Liner:Premium 4 mm Precision liner w/wool (cat.# 23305.5)
Inj. Temp.:200 °C
Oven
Oven Temp.:40 °C (hold 3 min) to 130 °C at 8 °C/min
Carrier GasHe, constant flow
Flow Rate:1.0 mL/min
DetectorMS
Mode:Scan
Transfer Line Temp.:200 °C
Analyzer Type:Quadrupole
Source Temp.:200 °C
Quad Temp.:150 °C
Electron Energy:70 eV
Solvent Delay Time:2.5 min
Tune Type:PFTBA
Ionization Mode:EI
InstrumentAgilent 7890A GC & 5975C MSD
NotesScan Program: 30-150 amu
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Wax

,

Stabilwax

,

solvents

,

glycols

,

glycol

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)
Drag and Drop Attachments Here (optional)
...


all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us