Commercial Spearmint Oil on Stabilwax-MS
This chromatogram may contain analytes that are synonyms of your search terms.

Commercial Spearmint Oil on Stabilwax-MS

PeakstR (min)
1.α-Pinene4.66
2.β-Pinene6.92
3.Sabinene7.37
4.Myrcene8.68
5.Limonene9.49
6.1,8-Cineole9.67
7.Cymene11.33
8.3-Octanol14.10
9.1-Octen-3-ol15.26
10.Menthone15.59
11.Menthofuran15.94
12.β-Bourbonene16.58
PeakstR (min)
13.Linalool17.11
14.Menthyl acetate17.44
15.4-Terpineol18.12
16.Dihydrocarvone18.26
17.Neoisomenthol18.59
18.Menthol*18.88
19.Terpineol19.76
20.Germacrene D20.01
21.Piperitone20.31
22.Carvone20.44
23.Jasmone23.66
24.Caryophyllene oxide24.28
* High signal suggests adulteration or misbranding.
image
GC_GN1169
ColumnStabilwax-MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 10673)
SampleCommercial spearmint oil
Conc.: Neat
Injection
Inj. Vol.:1 µL split (split ratio 150:1)
Liner:Premium 3.5 mm ID single taper w/wool (cat.# 23322.5)
Inj. Temp.:250 °C
Oven
Oven Temp.:45 °C (hold 5 min) to 250 °C at 7 °C/min (hold 10 min)
Carrier GasHe, constant linear velocity
Linear Velocity:36 cm/sec
DetectorMS
Mode:Scan
Scan Program:
GroupStart Time
(min)
Scan Range
(amu)
Scan Rate
(scans/sec)
1040-5503.3
Transfer Line Temp.:260 °C
Analyzer Type:Quadrupole
Source Temp.:250 °C
Electron Energy:70 eV
Ionization Mode:EI
InstrumentShimadzu 2010 GC & QP2010+ MS
Download PDF

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us