Comparison of Low-Level Pesticide Responses Using Restek Premium vs. Standard Deactivation Inlet Liners, Part I (QuEChERS Performance Standard C on Rxi-5ms)
This chromatogram may contain analytes that are synonyms of your search terms.

Comparison of Low-Level Pesticide Responses Using Restek Premium vs. Standard Deactivation Inlet Liners, Part I (QuEChERS Performance Standard C on Rxi-5ms)

Peaks
1.Vinclozoline
2.Carbaryl
3.Metalaxyl
4.Methiocarb
5.Cyprodinil
6.Thiabendazole
7.Captan
8.Folpet
image
GC_GN1185
ColumnRxi-5ms, 15 m, 0.25 mm ID, 0.25 µm (cat.# 13420)
SampleQuEChERS performance standard C (cat.# 31155)
Diluent:Toluene
Conc.:100 pg/µL
Injection
Inj. Vol.:1 µL splitless (hold 1 min)
Liner:Comparison of low-level pesticide responses using Premium single taper with wool (cat.# 23303.5) and standard deactivation single taper with wool (cat.# 22406)
Inj. Temp.:250 °C
Purge Flow:40 mL/min
Oven
Oven Temp.:70 °C (hold 1 min) to 330 °C at 21 °C/min (hold 1.5 min)
Carrier GasHe, constant flow
Flow Rate:1.5 mL/min
DetectorFID @ 340 °C
Make-up Gas Flow Rate:45 mL/min
Make-up Gas Type:N2
Hydrogen flow:40 mL/min
Air flow:400 mL/min
Data Rate:50 Hz
InstrumentAgilent 7890A GC
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Deactivation

,

Intermediate Polarity

,

Liner

,

Carbamates

,

Quartz Wool

,

Borosilicate Wool

,

QuEChERS Performance Standard C

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)
Drag and Drop Attachments Here (optional)
...


all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us