share:
delim
delim
Residual Solvents and Gases on Rxi-624Sil MS via HS-Syringe-GC-FID

Residual Solvents and Gases on Rxi-624Sil MS via HS-Syringe-GC-FID

PeakstR (min)Conc.
(µg/mL)
1.n-Propane1.49210
2.Isobutane1.70810
3.n-Butane1.90110
4.Methanol2.21510
5.n-Pentane3.01810
6.Ethanol3.21210
7.Diethyl ether3.34410
8.Acetone3.82010
9.2-Propanol4.18810
10.Acetonitrile4.36710
11.Methylene chloride4.71010
12.n-Hexane6.03310
13.Ethyl acetate7.67510
14.Chloroform8.10410
15.Cyclohexane8.32810
16.Benzene8.76310
17.1,2-Dichloroethane8.87210
18.n-Heptane9.23610
19.Trichloroethene9.63710
20.Toluene11.50010
21.m-Xylene13.42110
22.p-Xylene13.42110
23.o-Xylene13.72510
24.N,N-Dimethylacetamide (solvent)13.958
image
GC_GN1198
ColumnRxi-624Sil MS, 30 m, 0.25 mm ID, 1.40 µm (cat.# 13868)
SampleResidual solvents #1 (cat.# 34105)
n-Propane, isobutane, n-butane (Emerald Scientific)
Diluent:N,N-Dimethylacetamide
Conc.:10 µg/mL
Injection
Inj. Vol.:250 µL headspace-syringe split (split ratio 10:1)
Liner:Topaz 1.8 mm ID straight/SPME inlet liner (cat.# 23280)
Inj. Temp.:280 °C
Headspace-Syringe
Split Vent Flow Rate:20 mL/min
Instrument:PAL RTC
Syringe Temp.:150 °C
Sample Temp.:80 °C
Sample Equil. Time:45 min
Inj. Speed:1700 µL/sec
Oven
Oven Temp.:30 °C (hold 6 min) to 85 °C at 15 °C/min (hold 2 min) to 250 °C at 35 °C/min
Carrier GasHe, constant flow
Flow Rate:2 mL/min
Linear Velocity:40.63 cm/sec
DetectorFID @ 320 °C
Make-up Gas Flow Rate:40 mL/min
Make-up Gas Type:N2
Hydrogen flow:45 mL/min
Air flow:450 mL/min
Data Rate:20 Hz
InstrumentAgilent 7890B GC
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Analysis

,

Cannabis

,

Residual

,

Solvents

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us