share:
delim
delim
GC Amenable Pesticides Regulated by California in Gummy Edibles on Rxi-5ms

GC Amenable Pesticides Regulated by California in Gummy Edibles on Rxi-5ms

PeakstR (min)PolarityPrecursor IonProduct IonTransition Type
1.Atrazine-d56.82Positive220.058.0Quantifier
2.Atrazine-d56.82Positive205.0127.0Qualifier
3.Diazinon-d10 (diethyl-d10)7.01Positive183.0139.0Quantifier
4.Diazinon-d10 (diethyl-d10)7.01Positive183.0168.0Qualifier
5.Quintozene7.03Positive294.9236.9Quantifier
6.Quintozene7.03Positive236.8118.9Qualifier
7.Methyl parathion7.50Positive263.0109.0Quantifier
8.Methyl parathion7.50Positive263.079.0Qualifier
9.Captan8.37Positive184.0149.1Quantifier
10.Captan8.37Positive184.0134.1Qualifier
11.trans-Chlordane8.41Positive271.9237.0Quantifier
12.trans-Chlordane8.41Positive372.9265.9Qualifier
13.cis-Chlordane8.53Positive372.9265.9Quantifier
14.cis-Chlordane8.53Positive271.9237.0Qualifier
15.Chlorfenapyr8.80Positive247.1227.1Quantifier
16.Chlorfenapyr8.80Positive59.131.1Qualifier
17.Cyfluthrin10.61Positive226.0206.0Quantifier
18.Cyfluthrin10.61Positive163.0127.0Qualifier
19.Cypermethrin10.87Positive163.0127.1Quantifier
20.Cypermethrin10.87Positive181.1152.1Qualifier
image
GC_GN1207
ColumnRxi-5ms, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13423)
SampleCalifornia pesticide standard #1 (cat.# 34124)
California pesticide standard #2 (cat.# 34125)
California pesticide standard #3 (cat.# 34126)
California pesticide standard #4 (cat.# 34127)
California pesticide standard #5 (cat.# 34128)
California pesticide standard #6 (cat.# 34129)
Atrazine-d5 (cat.# 31984)
Diazinon-d10 (cat.# 31986)
Diluent:Acetonitrile
Conc.:2.5-10 ng/mL Expected concentration range in extract after extracting from gummy fortified at 100 ng/g (final extract was diluted in half with acetonitrile).
Injection
Inj. Vol.:1 µL splitless
Liner:Topaz 4.0 mm ID single taper inlet liner w/wool (cat.# 23447)
Inj. Temp.:250 °C
Purge Flow:5 mL/min
Oven
Oven Temp.:90 °C (hold 1 min) to 310 °C at 25 °C/min (hold 10 min)
Carrier GasHe, constant flow
Flow Rate:1.4 mL/min
DetectorMS/MS
Transfer Line Temp.:290 °C
Analyzer Type:Quadrupole
Source Temp.:330 °C
Electron Energy:70 eV
Tune Type:PFTBA
Ionization Mode:EI
InstrumentThermo Scientific TSQ 8000 Triple Quadrupole GC-MS
NotesGummies were manually chopped into small pieces, and 1 g of sample was weighed in a 50 mL polypropylene tube. The sample was mixed with 5 mL of water and then vigorously vortexed until all gummy pieces were fully solubilized. The sample was fortified with pesticides and mycotoxins at 100 ng/g. A mix of internal standards was added at 200 ng/g. The spiked sample was further vortexed for 30 sec. 5 mL of acetonitrile acidified with 1% acetic acid was added to the sample, and this was followed by 30 sec vortex agitation. Then, a pouch of European EN 15662 QuEChERS extraction salts (cat.# 25849) was added to the sample. The sample was vortexed for 30 sec and then centrifuged for 5 min. 1.9 mL of supernatant was transferred to a Q-sep QuEChERS dSPE tube containing pre-weighed magnesium sulfate, PSA, and GCB (cat.# 26217). After vortexing and centrifuging, 500 μL of extract was mixed with 500 μL of acidified acetonitrile. 1 μL of final extract was injected into the GC-MS/MS system.


Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Pesticides

,

,

California

,

Cannabis

,

QuEChERS

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)
Drag and Drop Attachments Here (optional)
...


all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us