share:
delim
delim
Unleaded Gasoline on Rtx-1 (Comparison of Unweathered and 99% Weathered)
This chromatogram may contain analytes that are synonyms of your search terms.

Unleaded Gasoline on Rtx-1 (Comparison of Unweathered and 99% Weathered)

image
GC_MS00270
ColumnRtx-1, 30 m, 0.53 mm ID, 1.50 µm (cat.# 10170)
Injectionsplit (split ratio 30:1)
Inj. Temp.:250 °C
Oven
Oven Temp.:40 °C (hold 3 min) to 75 °C at 15 °C/min to 275 °C at 20 °C/min (hold 5 min)
Carrier GasH2, constant flow
Linear Velocity:50 cm/sec @ 40 °C
DetectorFID @ 285 °C
NotesFID sensitivity: 4.10 x 10-9 AFS
Download PDF

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us