share:
delim
delim
Permanent Gases Rt-Q-BOND

Permanent Gases Rt-Q-BOND

Peaks
1.Helium
2.Air
3.Carbon monoxide
4.Methane
5.Carbon dioxide
image
GC_PC00169
ColumnRt-Q-BOND, 30 m, 0.32 mm ID, 10 µm (cat.# 19744)
Sample2-5 mole %
Injection
Inj. Vol.:30 µL split (split ratio 40:1)
Inj. Temp.:30 °C
Oven
Oven Temp.:30 °C
Carrier GasH2, constant flow
Linear Velocity:38 cm/sec
DetectorHP-μTCD @ 200 °C
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Rt-QPLOT™

,

PLOT

,

permanent gas

,

gas analysis

,

Permanent

,

Gases

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us