share:
delim
delim
Aromatics (Benzene/Toluene/Xylene) on MXT-WAX
This chromatogram may contain analytes that are synonyms of your search terms.

Aromatics (Benzene/Toluene/Xylene) on MXT-WAX

Peaks
1.Benzene
2.Toluene
3.Ethylbenzene
4.p-Xylene
5.m-Xylene
6.o-Xylene
image
GC_PC00188
ColumnMXT-WAX, 30 m, 0.28 mm ID, 1.00 µm (cat.# 70654)
SampleBTEX Standard (cat.# 30051)
Injection
Inj. Vol.:1.0 µL split (split ratio 100:1)
Inj. Temp.:260 °C
Oven
Oven Temp.:60 °C
Carrier GasH2, constant flow
Linear Velocity:38 cm/sec
DetectorFID @ 260 °C
NotesFID sensitivity: 3.2 x 10-11 AFS
Download PDF

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us