share:
delim
delim
Aromatics (Benzene/Toluene/Xylene) on Rtx-502.2
This chromatogram may contain analytes that are synonyms of your search terms.

Aromatics (Benzene/Toluene/Xylene) on Rtx-502.2

Peaks
1.Methyl tert-butyl ether
2.Benzene
3.Toluene
Peaks
4.Ethylbenzene
5.m/p-Xylene
6.o-Xylene
image
GC_PC00190
ColumnRtx-502.2, 105 m, 0.53 mm ID, 3.00 µm (cat.# 10910)
SampleCA PVOC Mix (cat.# 30071)
Conc.:1000-5000 µg/mL
Injection
Inj. Vol.:0.25 µL split
Inj. Temp.:200 °C
Split Vent Flow Rate:70 mL/min
Oven
Oven Temp.:40 °C (hold 1 min) to 100 °C at 5 °C/min to 240 °C at 8 °C/min (hold 8 min)
Carrier GasHe, constant flow
Linear Velocity:67 cm/sec @ 40 °C
DetectorFID @ 280 °C
NotesFID sensitivity: 2.05 x 10-9 AFS