Sim Dist, ASTM D3710 (Gasoline) on Rtx-1 SimDist 2887
This chromatogram may contain analytes that are synonyms of your search terms.

Sim Dist, ASTM D3710 (Gasoline) on Rtx-1 SimDist 2887

Peaks
1.n-Propane
2.2-Methyl propane
3.n-Butane
4.2-Methylbutane
5.n-Pentane
6.2-Methylpentane
7.n-Hexane
8.2,4-Dimethylpentane
9.n-Heptane
10.Toluene
11.p-Xylene
12.n-Propylbenzene
13.n-Decane
14.n-Butylbenzene
image
GC_PC00230
ColumnRtx-1 SimDist 2887, SilcoSmooth Tubing, 25 inches, 1/8 in. OD, 2 mm ID (cat.# 80000-800)
Sampleunleaded gasoline (ASTM D2887)
Injection
Inj. Vol.:1.0 µL packed not on-column
Inj. Temp.:250 °C
Oven
Oven Temp.:-30 °C to 250 °C at 10 °C/min
Carrier GasHe, constant flow
Flow Rate:25 mL/min
DetectorFID @ 300 °C
NotesFID sensitivity: 256 x 10-11 AFS