Hydrocarbons (C10-C44) on MXT-1HT SimDist

GC_PC00529
Column | MXT-1HT SimDist, 5 m, 0.53 mm ID, 0.10 µm (cat.# 70100) |
Sample | D2887 calibration mix (cat.# 31222) |
Diluent: | Carbon disulfide |
Injection | |
Inj. Vol.: | 0.2 µL on-column |
Inj. Temp.: | Track oven |
Oven | |
Oven Temp.: | 40 °C to 430 °C at 10 °C/min (hold 30 min) |
Carrier Gas | He, constant pressure (1.0 psi, 6.9 kPa) |
Linear Velocity: | 60 cm/sec |
Detector | FID @ 430 °C |
Make-up Gas Flow Rate: | 40 mL/min |
Notes | Flow rate: 7.8 mL/min |