share:
delim
delim
Derivatized B100 and Internal Standards on MXT-Biodiesel TG with 2m Integra Gap

Derivatized B100 and Internal Standards on MXT-Biodiesel TG with 2m Integra Gap

image
GC_PC00934
ColumnMXT-Biodiesel TG 13m, 0.53mm ID, 0.16µm with built-in 2m Integra-Gap (total column length 15 m)
SampleB100 + IS Butanetriol & Tricaprin derivatized with MSTFA as per ASTM D-6584
Injection
Inj. Vol.:1.0 µL cold on-column
Temp. Program:oven track
Oven
Oven Temp.:50 °C (hold 1 min) to 180 °C at 15 °C/min to 230 °C at 7 °C/min to 430 °C at 30 °C/min (hold 5 min)
Carrier GasH2, constant flow
Flow Rate:4 mL/min
DetectorFID @ 430 °C
InstrumentShimadzu 2010 GC
Notes(Data acquired on prototype column, similar column to cat# 70290)

RELATED CHROMATOGRAM LIBRARY SEARCHES

total glycerides

,

ASTM D-6584

,

EN14105

,

residual triglyceride

,

glycerides

,

biodiesel

,

B100

,

ASTM

,

6584

,

triglycerides

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us