Gasoline Containing Ethanol by ASTM D3606-06 (Modified) on D3606 Application Column Set
This chromatogram may contain analytes that are synonyms of your search terms.

Gasoline Containing Ethanol by ASTM D3606-06 (Modified) on D3606 Application Column Set

Peaks
1.Ethanol
2.Benzene
3.2-Butanol (IS)
4.Toluene
image
GC_PC01079
ColumnD3606 Application Column (2 column set) Column 1: 6' (1.8m), 1/8" OD, 2.0mm ID, nonpolar Rtx®-1 Column 2: 16' (4.9m), 1/8" OD, 2.0mm ID, proprietary packing material (cat.# 83606-800)
SampleGasoline with internal standard
Diluent:
Injectionsample valve
Sample Loop Vol.:1.5 µL
Oven
Oven Temp.:135 °C (hold 16 min)
Carrier GasHe, constant flow
Flow Rate:20.0 mL/min
DetectorTCD @ 200 °C
Notes3 min. backflush
AcknowledgementChromatogram courtesy of Boguslaw Dudek, Conoco Phillips, Linden, NJ