share:
delim
delim
Refinery Gasoline Sample on D3606 Application Column Set

Refinery Gasoline Sample on D3606 Application Column Set

Peaks
1.Ethanol
2.Benzene
3.2-Butanol (IS)
4.Toluene
image
GC_PC01106
ColumnD3606 Application Column (2 column set) Column 1: 6' (1.8m), 1/8" OD, 2.0mm ID, nonpolar Rtx-1 Column 2: 16' (4.9m), 1/8" OD, 2.0mm ID, proprietary packing material (cat.# 83606-810)
Samplerefinery gasoline sample
Injectionsample valve
Sample Loop Vol.:1.5 µL
Oven
Oven Temp.:135 °C
Carrier GasHe, constant flow
Flow Rate:20 mL/min
DetectorTCD @ 200 °C
InstrumentAgilent/HP6890 GC
NotesBackflush time: 3 min
Inj. temp.: 200°C
AcknowledgementCourtesy of Bogey Dudek, ConocoPhillips, US
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

D3606

,

gasoline

,

3606

,

astm

,

refinery gas

,

packed column

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us