Refinery Gasoline Sample on D3606 Application Column Set
This chromatogram may contain analytes that are synonyms of your search terms.

Refinery Gasoline Sample on D3606 Application Column Set

Peaks
1.Ethanol
2.Benzene
3.2-Butanol (IS)
4.Toluene
image
GC_PC01106
ColumnD3606 Application Column (2 column set) Column 1: 6' (1.8m), 1/8" OD, 2.0mm ID, nonpolar Rtx-1 Column 2: 16' (4.9m), 1/8" OD, 2.0mm ID, proprietary packing material (cat.# 83606-810)
Samplerefinery gasoline sample
Injectionsample valve
Sample Loop Vol.:1.5 µL
Oven
Oven Temp.:135 °C
Carrier GasHe, constant flow
Flow Rate:20 mL/min
DetectorTCD @ 200 °C
InstrumentAgilent/HP6890 GC
NotesBackflush time: 3 min
Inj. temp.: 200°C
AcknowledgementCourtesy of Bogey Dudek, ConocoPhillips, US