share:
delim
delim
Ethylene Purity Analysis using ASTM D6159-97

Ethylene Purity Analysis using ASTM D6159-97

Peaks
1.Methane
2.Ethane
3.Ethylene
image
GC_PC01109
Columnsee notes
Sampleethylene gas
Injection
Inj. Vol.:1.0 µL split
Liner:2mm Splitless (cat.# 20712)
Inj. Temp.:200 °C
Split Vent Flow Rate:60 mL/min
Oven
Oven Temp.:35 °C (hold 2 min) to 190 °C at 4 °C/min (hold 15 min)
Carrier GasHe, constant pressure (8.0 psi, 55.2 kPa)
Linear Velocity:25.4 cm/sec @ 35 °C
DetectorFID @ 200 °C
Make-up Gas Type:N2
Data Rate:20 Hz
InstrumentHP5890 GC
NotesRt-Alumina BOND/KCl, 50m, 0.53mm ID, 10.0um (cat. # 19760) in series with Rtx-1, 30m, 0.53mm ID, 5.0um (cat. #10179), connected with a Universal Press-Tight Connector (cat. # 20401)
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

ASTM D6159-97

,

Ethylene

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us