Gasoline on ASTM D3606 Application Column Set
This chromatogram may contain analytes that are synonyms of your search terms.

Gasoline on ASTM D3606 Application Column Set

Peaks
1.Ethanol
2.Benzene
3.2-Butanol (IS)
4.Toluene
image
GC_PC01121
ColumnD3606 Application Column (2 column set) Column 1: 6' (1.8m), 1/8" OD, 2.0mm ID, nonpolar Rtx®-1; Column 2: 16' (4.9m), 1/8" OD, 2.0mm ID, proprietary packing material (cat.# 83606-800)
Samplecommercial gasoline
Injection
Inj. Vol.:1 µL Direct
Inj. Temp.:200 °C
Oven
Oven Temp.:135 °C
Carrier GasHe, constant flow
Flow Rate:20 mL/min
DetectorTCD @ 230 °C
InstrumentAgilent/HP6890 GC
NotesValve box temp.: 200°C
Backflush time: 3 min.
Download PDF