Refinery Gas Hydrocarbons on Rt-Alumina BOND/Na2SO4
This chromatogram may contain analytes that are synonyms of your search terms.

Refinery Gas Hydrocarbons on Rt-Alumina BOND/Na2SO4

Peaks
1.Methane
2.Ethane
3.Ethylene
4.Propane
5.Propylene
6.Isobutane
7.n-Butane
8.Propadiene
9.Acetylene
10.trans-2-Butene
Peaks
11.1-Butene
12.Isobutylene
13.cis-2-Butene
14.Isopentane
15.n-Pentane
16.1,3-Butadiene
17.trans-2-Pentene
18.2-Methyl-2-butene
19.1-Pentene
20.cis-2-Pentene
21.n-Hexane
image
GC_PC01139
ColumnRt-Alumina BOND/Na2SO4, 30 m, 0.53 mm ID, 10 µm (cat.# 19755)
SampleRefinery gas hydrocarbons through C6
Injection
Inj. Vol.:10 µL split
Liner:2 mm split Precision liner w/wool (cat.# 20823)
Inj. Temp.:200 °C
Split Vent Flow Rate:40 mL/min
Oven
Oven Temp.:60 °C (hold 2 min) to 200 °C at 10 °C/min (hold 1 min)
Carrier GasHe, constant pressure (5.0 psi, 34.5 kPa)
Linear Velocity:37.3 cm/sec @ 60 °C
DetectorFID @ 200 °C
Make-up Gas Type:N2
Data Rate:20 Hz
InstrumentHP5890 GC
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Rt-Alumina Bond/Na2SO4

,

Refinery Gas

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us