share:
delim
delim
Propylene and C1-C5 Hydrocarbons on Rt-Alumina BOND/Na2SO4
This chromatogram may contain analytes that are synonyms of your search terms.

Propylene and C1-C5 Hydrocarbons on Rt-Alumina BOND/Na2SO4

Peaks
1.Methane
2.Ethane
3.Ethylene
4.Propane
5.Cyclopropane
6.Propylene
7.Isobutane
8.n-Butane
9.Propadiene
Peaks
10.Acetylene
11.trans-2-Butene
12.1-Butene
13.Isobutylene
14.cis-2-Butene
15.Isopentane
16.n-Pentane
17.1,3-Butadiene
18.Methylacetylene
image
GC_PC01142
ColumnRt-Alumina BOND/Na2SO4, 50 m, 0.53 mm ID, 10 µm (cat.# 19756)
SamplePropylene and C1-C5 hydrocarbons
Injection
Inj. Vol.:10 µL split
Liner:2 mm splitless (cat.# 20712)
Inj. Temp.:200 °C
Split Vent Flow Rate:35 mL/min
Oven
Oven Temp.:50 °C to 150 °C at 6 °C/min
Carrier GasHe, constant pressure (10 psi, 68.9 kPa)
Linear Velocity:47 cm/sec @ 50 °C
DetectorFID @ 200 °C
Make-up Gas Type:N2
Data Rate:20 Hz
InstrumentHP5890 GC
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Propylene

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)
Drag and Drop Attachments Here (optional)
...


all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us