share:
delim
delim
Gasoline Light Fractions on Rtx-DHA-150

Gasoline Light Fractions on Rtx-DHA-150

Peaks
1.Methane
2.Ethane
3.Propylene
4.Propane
5.Isobutane
6.Butane
image
GC_PC1145
ColumnRtx-DHA-150, 150 m, 0.25 mm ID, 1.0 µm (cat.# 10149)
SampleGasoline
Conc.: Neat
Injection
Inj. Vol.:1.0 µL split (split ratio 150:1)
Liner:Cyclosplitter (cat.# 20706)
Inj. Temp.:275 °C
Oven
Oven Temp.:35 °C (hold 60 min) to 200 °C at 5 °C/min (hold 5 min)
Carrier GasH2, constant flow
Linear Velocity:44 cm/sec
DetectorFID @ 275 °C
Make-up Gas Flow Rate:30 mL/min
Make-up Gas Type:N2
InstrumentAgilent/HP6890 GC
NotesLight fractions of gasoline above ambient temperature
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

light fractions of gasoline

,

light fractions

,

DHA-150

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us