Low Bleed, High Efficiency MXT-1HT SimDist Columns Outperform Competitors for Simulated Distillation (ASTM D6352)

GC_PC1149
Column | Restek MXT-1HT SimDist (cat.#70115), see above for competitor column descriptions |
Sample | C8-C100 hydrocarbons |
Diluent: | CS2 |
Conc.: | 1.0% |
Injection | |
Inj. Vol.: | 0.5 µL cold on-column |
Temp. Program: | 53 °C to 430 °C at 10 °C/min (hold 5 min) |
Oven | |
Oven Temp.: | 50 °C to 430 °C at 10 °C/min (hold 5 min) |
Carrier Gas | He, constant flow |
Flow Rate: | 18 mL/min |
Detector | FID @ 430 °C |
Make-up Gas Flow Rate: | 24 mL/min |
Make-up Gas Type: | N2 |
Instrument | Shimadzu 2010 GC |