Pure Chromatography
www.restek.com
www.restek.com

This chromatogram may contain analytes that are synonyms of your search terms.
C10-C40 Alkanes on Rxi-5HT
Peaks | |
---|---|
1. | C10![]() |
2. | C12![]() |
3. | C14![]() |
4. | C16![]() |
5. | C18![]() |
6. | C20![]() |
7. | C22![]() |
8. | C24![]() |
Peaks | |
---|---|
9. | C26![]() |
10. | C28![]() |
11. | C30![]() |
12. | C32![]() |
13. | C34![]() |
14. | C36![]() |
15. | C38![]() |
16. | C40![]() |

GC_PC1159
Column | Rxi-5HT, 30 m, 0.32 mm ID, 0.25 µm (cat.# 13924) |
using Rxi guard column 2 m, 0.53 mm ID (cat.# 10054) | |
Sample | System performance test standard mixture of n-alkanes (cat.# 31678) |
Diluent: | Hexane |
Conc.: | 50 µg/mL |
Injection | |
Inj. Vol.: | 1 µL cold on-column |
Temp. Program: | 53 °C to 300 °C at 10 °C/min (hold 20 min) |
Oven | |
Oven Temp.: | 50 °C to 300 °C at 10 °C/min (hold 20 min) |
Carrier Gas | H2, constant flow |
Linear Velocity: | 40 cm/sec @ 50 °C |
Dead Time: | 1.25 min @ 50 °C |
Detector | FID @ 330 °C |
Make-up Gas Flow Rate: | 30 mL/min |
Make-up Gas Type: | N2 |
Data Rate: | 20 Hz |
Instrument | Agilent/HP6890 GC |
RELATED CHROMATOGRAM LIBRARY SEARCHES
Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.