share:
delim
delim
Diesel on Rtx-Mineral Oil

Diesel on Rtx-Mineral Oil

image
GC_PC1171
ColumnRtx-Mineral Oil, 15 m, 0.32 mm ID, 0.15 µm (cat.# 18074)
using IP Deactivated Guard Column 2 m, 0.53 mm ID (cat.# 10047)
Samplediesel fuel #2 composite (cat.# 31093)
Diluent:methylene chloride
Conc.:5,000 µg/mL diluted to 500 μg/mL
Injection
Inj. Vol.:0.5 µL cold on-column
Temp. Program:53 °C to 300 °C at 10 °C/min (hold 20 min)
Oven
Oven Temp.:50 °C to 300 °C at 10 °C/min (hold 20 min)
Carrier GasH2, constant flow
Linear Velocity:40 cm/sec @ 50 °C
Dead Time:0.625 min @ 50 °C
DetectorFID @ 330 °C
Make-up Gas Flow Rate:30 mL/min
Make-up Gas Type:N2
Data Rate:20 Hz
InstrumentAgilent/HP6890 GC
NotesBlack trace = diesel sample
Red trace = C10-C30 standard
Download PDF

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us