share:
delim
delim
C5-C72 Hydrocarbons on MXT-1HT SimDist by ASTM D7213-05 (D2887 Extended))
This chromatogram may contain analytes that are synonyms of your search terms.

C5-C72 Hydrocarbons on MXT-1HT SimDist by ASTM D7213-05 (D2887 Extended))

image
GC_PC1191
ColumnMXT-1HT SimDist, 5 m, 0.53 mm ID, 0.10 µm (cat.# 70112)
SampleC5-C72 hydrocarbons (Separation Systems SD-ss3e-03)
Diluent:CS2
Conc.:1.0%
Injection
Inj. Vol.:0.5 µL cold on-column
Temp. Program:53 °C to 380 °C at 10 °C/min (hold 12 min)
Oven
Oven Temp.:50 °C to 380 °C at 10 °C/min (hold 12 min)
Carrier GasHe, constant flow
Flow Rate:12 mL/min
DetectorFID @ 390 °C
Make-up Gas Flow Rate:18 mL/min
Make-up Gas Type:N2
InstrumentShimadzu 2010 GC
Download PDF

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)
Drag and Drop Attachments Here (optional)
...


all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us