share:
delim
delim
Gasoline Containing Ethanol on D3606 Application Column Set by ASTM D3606-10 (Modified)

Gasoline Containing Ethanol on D3606 Application Column Set by ASTM D3606-10 (Modified)

Peaks
1.Ethanol
2.Benzene
3.2-Butanol (IS)
4.Toluene
image
GC_PC1224
ColumnD3606 application column (2 column set). Column 1: 6' (1.8 m), 1/8" OD, 2.0 mm ID, nonpolar Rtx®-1; Column 2: 16' (4.9 m), 1/8" OD, 2.0 mm ID, proprietary packing material (cat.# 83606-800)
SampleEthanol-containing gasoline with internal standard (IS)
Diluent:
Injection
Inj. Vol.:1.5 µL Direct
Inj. Temp.:150 °C
Oven
Oven Temp.:135 °C (hold 12 min)
Carrier GasHe, constant flow
Flow Rate:20.0 mL/min
DetectorTCD @ 200 °C
Notes2.38 minute backflush (must be determined for each GC system).
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

ethanol

,

gasoline

,

D3606

,

ASTM

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us