share:
delim
delim
Stability Overlay: 1st and 40th Injections of Short-Chain Amines in Water on Rtx-Volatile Amine

Stability Overlay: 1st and 40th Injections of Short-Chain Amines in Water on Rtx-Volatile Amine

Peaks
1.Methanol
2.Dimethylamine
3.Trimethylamine
4.Methylethylamine
5.Dimethylethylamine
6.Diethylamine
7.Methyldiethylamine
8.Triethylamine
image
GC_PC1244
ColumnRtx-Volatile Amine, 60 m, 0.32 mm ID (cat.# 18078)
SampleShort-chain amines
Diluent:Water
Injection
Inj. Vol.:1 µL split (split ratio 15:1)
Inj. Temp.:220 °C
Split Vent Flow Rate:30 mL/min
Oven
Oven Temp.:40 °C (hold 10 min) to 250 °C at 20 °C/min (hold 10 min)
Carrier GasH2, constant flow
Flow Rate:2 mL/min
Linear Velocity:35 cm/sec
DetectorFID @ 250 °C
InstrumentAgilent/HP6890 GC
AcknowledgementCourtesy of Gilbert Baele, Taminco (Antwerp, Belgium)
Download PDF

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us