share:
delim
delim
Impurities in Methyl-t-Butyl Ether (MTBE) on Rtx-DHA-100 (ASTM D5441)
This chromatogram may contain analytes that are synonyms of your search terms.

Impurities in Methyl-t-Butyl Ether (MTBE) on Rtx-DHA-100 (ASTM D5441)

Peaks
1.Methanol
2.iso-Pentane
3.n-Pentane
4.trans-2-Pentene
5.tert-Butanol
6.cis-2-Pentene
Peaks
7.2-Methyl-2-butene
8.tert-Butyl ethyl ether
9.tert-Amyl methyl ether (TAME)
10.2,4,4-Trimethyl-1-pentene
11.4,4'-Dimethyl-2-neopentyl-1-pentene (triisobutylene)
12.2,2,4,6,6-Pentamethyl-3-heptene (triisobutylene)
image
GC_PC1268
ColumnRtx-DHA-100, 100 m, 0.25 mm ID, 0.50 µm (cat.# 10148)
SampleASTM D5441 MTBE contaminants mix
Conc.:1 % (w/w) each component in MTBE
Injection
Inj. Vol.:0.2 µL split (split ratio 200:1)
Liner:Cup splitter (cat.# 20710)
Inj. Temp.:200 °C
Oven
Oven Temp.:50 °C (hold 13 min) to 180 °C at 10 °C/min (hold 7 min)
Carrier GasHe, constant pressure (39.3 psi, 271.0 kPa)
Linear Velocity:23 cm/sec @ 50 °C
DetectorFID @ 250 °C
Make-up Gas Flow Rate:24 mL/min
Make-up Gas Type:N2
Hydrogen flow:30 mL/min
Air flow:358 mL/min
InstrumentAgilent 7890A GC
NotesResolution between trans-2-pentene/tert-butanol and tert-butanol/cis-2-pentene exceeds the ASTM Method D5441 requirement of R ≥ 1.3.
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

D5441

,

MTBE impurities

,

ASTM D5441

,

methyl tert-butyl ether

,

ASTM

,

,

DHA

,

gasoline

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us